Back to Search Start Over

Automated Off-Line Respiratory Event Detection for the Study of Postoperative Apnea in Infants

Authors :
Karen A. Brown
Ahmed Aoude
Carlos A. Robles-Rubio
Robert E. Kearney
Henrietta L. Galiana
Source :
IEEE Transactions on Biomedical Engineering. 58:1724-1733
Publication Year :
2011
Publisher :
Institute of Electrical and Electronics Engineers (IEEE), 2011.

Abstract

Previously, we presented automated methods for thoraco-abdominal asynchrony estimation and movement artifact detection in respiratory inductance plethysmography (RIP) signals. This paper combines and improves these methods to give a method for the automated, off-line detection of pause, movement artifact, and asynchrony. Simulation studies demonstrated that the new combined method is accurate and robust in the presence of noise. The new procedure was successfully applied to cardiorespiratory signals acquired postoperatively from infants in the recovery room. A comparison of the events detected with the automated method to those visually scored by an expert clinician demonstrated a higher agreement (κ = 0.52) than that amongst several human scorers (κ = 0.31) in a clinical study . The method provides the following advantages: first, it is fully automated; second, it is more efficient than visual scoring; third, the analysis is repeatable and standardized; fourth, it provides greater agreement with an expert scorer compared to the agreement between trained scorers; fifth, it is amenable to online detection; and lastly, it is applicable to uncalibrated RIP signals. Examples of applications include respiratory monitoring of postsurgical patients and sleep studies.

Details

ISSN :
15582531 and 00189294
Volume :
58
Database :
OpenAIRE
Journal :
IEEE Transactions on Biomedical Engineering
Accession number :
edsair.doi.dedup.....28e025a7b126f7c6e6e9c828ec573aeb
Full Text :
https://doi.org/10.1109/tbme.2011.2112657