Back to Search Start Over

Reel syndrome in an ICD with DF4 connector

Authors :
Christian G. Wollmann
Source :
Herzschrittmachertherapie + Elektrophysiologie. 25:206-208
Publication Year :
2014
Publisher :
Springer Science and Business Media LLC, 2014.

Abstract

This is the first case report about a patient implanted with a DF4 lead suffering from reel syndrome. The reel syndrome may occur due to a ratchet mechanism where the device generator creates torque on a lead fixed at the fixation sleeve. Theoretically, this is less likely to occur in DF4 connectors due to their higher stiffness.