Back to Search Start Over

Condylar failure of the lacey rotating-hinge total knee

Authors :
John P. Heiner
F. J. Worzala
Christopher E. Scott
Ray Vanderby
Source :
The Journal of Arthroplasty. 11:214-216
Publication Year :
1996
Publisher :
Elsevier BV, 1996.

Abstract

The Lacey Rotating-Hinge (Wright Medical, Arlington, TN) total knee is a constrained device for reconstruction of knees without collateral ligaments or to replace resected bone. Problems have been described with hinged total knees such as stem fracture, particulate synovitis, and stem loosening. A different mode of failure is described here. Two case histories are provided in which similar fractures occurred in the lateral condyle of Lacey Rotating-Hinge total knees.

Details

ISSN :
08835403
Volume :
11
Database :
OpenAIRE
Journal :
The Journal of Arthroplasty
Accession number :
edsair.doi.dedup.....2d60154ea032195eafef7f03e91a72df
Full Text :
https://doi.org/10.1016/s0883-5403(05)80022-1