Back to Search
Start Over
Condylar failure of the lacey rotating-hinge total knee
- Source :
- The Journal of Arthroplasty. 11:214-216
- Publication Year :
- 1996
- Publisher :
- Elsevier BV, 1996.
-
Abstract
- The Lacey Rotating-Hinge (Wright Medical, Arlington, TN) total knee is a constrained device for reconstruction of knees without collateral ligaments or to replace resected bone. Problems have been described with hinged total knees such as stem fracture, particulate synovitis, and stem loosening. A different mode of failure is described here. Two case histories are provided in which similar fractures occurred in the lateral condyle of Lacey Rotating-Hinge total knees.
- Subjects :
- Male
Reoperation
musculoskeletal diseases
Orthodontics
business.industry
Total knee arthroplasty
Rotating hinge
Prosthesis Design
musculoskeletal system
Condyle
Total knee
Prosthesis Failure
Postoperative Complications
Humans
Medicine
Orthopedics and Sports Medicine
Range of Motion, Articular
Knee Prosthesis
business
human activities
Aged
Subjects
Details
- ISSN :
- 08835403
- Volume :
- 11
- Database :
- OpenAIRE
- Journal :
- The Journal of Arthroplasty
- Accession number :
- edsair.doi.dedup.....2d60154ea032195eafef7f03e91a72df
- Full Text :
- https://doi.org/10.1016/s0883-5403(05)80022-1