A novel and cheap method to correlate subjective and objective visual acuity by using the optokinetic response
- Source :
- International Ophthalmology. 38:2101-2115
- Publication Year :
- 2017
- Publisher :
- Springer Science and Business Media LLC, 2017.
-
Abstract
- To describe a novel optokinetic visual acuity estimator (Oktotype) and to report the preliminary results obtained in poorly and non-collaborative subjects. Eleven series of symbols arranged horizontally and moving from left to right at a constant rate were displayed. In each sequence, the size of the stimuli was reduced logarithmically. By using this paradigm, the objective visual acuity was computed in 26 normal subjects as the minimum size of the symbols able to evoke the optokinetic response. In the preliminary phase, three contrast levels were tested, with white noise added to the first five sequences so as to normalize the overestimate found at the lower-half range of the acuity scale. Subsequently, the correspondence between subjective and objective visual acuity was compared in 10 poorly collaborative subjects, and the agreement between optokinetic and Teller visual acuity was measured in six non-collaborative subjects. The best agreement is provided by the minimum contrast level (20%) (R 2 = 0.74). The correspondence between the two techniques is satisfying both in the normal and in the poorly collaborative sample (concordance correlation coefficient: 0.85 and 0.83, respectively). In the non-collaborative group, the concordance correlation coefficient between Teller acuity and OKVA ranged between 0.79 (test) and 0.85 (retest). Test–retest reliability was very good for the Oktotype (K: 0.82), and better than the Teller test (K = 0.71), even if it was lower compared to Snellen acuity (K = 0.95). The Oktotype seems promising to predict Snellen visual acuity in normal and poorly collaborative subjects.
- Subjects :
- Adult
Male
medicine.medical_specialty
Visual acuity
genetic structures
Cost-Benefit Analysis
media_common.quotation_subject
Visual Acuity
Audiology
Snellen acuity
Young Adult
03 medical and health sciences
0302 clinical medicine
Humans
Medicine
Contrast (vision)
Diagnosis, Computer-Assisted
Nystagmus, Optokinetic
media_common
Contrast level
business.industry
Vision Tests
Reproducibility of Results
Optokinetic reflex
Snellen visual acuity
Ophthalmology
Constant rate
Concordance correlation coefficient
030221 ophthalmology & optometry
Female
medicine.symptom
business
030217 neurology & neurosurgery
Subjects
Details
- ISSN :
- 15732630 and 01655701
- Volume :
- 38
- Database :
- OpenAIRE
- Journal :
- International Ophthalmology
- Accession number :