Interference of Asfotase Alfa in Immunoassays Employing Alkaline Phosphatase Technology
- Source :
- The Journal of Applied Laboratory Medicine. 5:290-299
- Publication Year :
- 2019
- Publisher :
- Oxford University Press (OUP), 2019.
-
Abstract
- Background Asfotase alfa (STRENSIQ®, Alexion Pharmaceuticals, Inc.) is the only approved treatment for patients with pediatric-onset hypophosphatasia, a disease caused by a mutation in the tissue-nonspecific alkaline phosphatase (TNSALP) gene. ALP is often used as signaling system in routine immunoassays. Because asfotase alfa contains the active site of the full ALP enzyme, it can catalyze the substrate as the antibody-conjugated ALP would within an assay. Therefore, its presence in a treated patient’s sample may generate false positive or false negative results. We investigated whether the presence of asfotase alfa within a sample induced interference in immunoassays that utilize ALP or alternative detection systems. Methods Asfotase alfa was added to samples at concentrations from 0.08–5 µg/mL and analysed on various immunoassays following manufacturer’s instructions. Results Asfotase alfa was detected in all ALP assays but ALKP1 (RayBiotech). We observed no changes in normetanephrine and noradrenaline (IBL) at any asfotase alfa concentration. However, asfotase alfa notably interfered in an oxytocin (ENZO) assay in nonextracted samples. Extraction using a C18 column eliminated the interference. No interference was observed on automated analyzers using alternative detection system (COBAS fT4 and TSH; Advia Centaur FSH, fT4; Architect LH; FSH). Immulite 2000 fT4, TSH, testosterone and hCG (ALP-based) showed no interference. However, the presence of asfotase alfa resulted in a dose-dependent increase of Troponin I signal. Conclusion The presence of asfotase alfa must be taken into consideration when analyzing blood samples in treated patients to avoid any risk of misinterpretation of false positive/negative results. It is essential that assays be tested for this possible interference.
- Subjects :
- 0301 basic medicine
030213 general clinical medicine
medicine.medical_specialty
Recombinant Fusion Proteins
Hypophosphatasia
Normetanephrine
03 medical and health sciences
chemistry.chemical_compound
0302 clinical medicine
stomatognathic system
Internal medicine
Troponin I
medicine
Humans
False Positive Reactions
ADVIA Centaur
Immunoassay
Analysis of Variance
medicine.diagnostic_test
Chemistry
General Medicine
Alkaline Phosphatase
medicine.disease
Signaling system
Enzyme Activation
030104 developmental biology
Endocrinology
Immunoglobulin G
Asfotase alfa
Alkaline phosphatase
Biomarkers
Subjects
Details
- ISSN :
- 24757241 and 25769456
- Volume :
- 5
- Database :
- OpenAIRE
- Journal :
- The Journal of Applied Laboratory Medicine
- Accession number :
- edsair.doi.dedup.....314c3ea1ef9865b4a74d2d0c65011ae5
- Full Text :
- https://doi.org/10.1093/jalm/jfz007