Comparison of valve performance of the mechanically expanding Lotus and the balloon-expanded SAPIEN3 transcatheter heart valves: an observational study with independent core laboratory analysis

Authors :
Hans Jonker
Gerrit Anne van Es
Osama Ibrahim Ibrahim Soliman
Lennart van Gils
Ben Ren
Jan G.P. Tijssen
Peter de Jaegere
Nicolas M. Van Mieghem
Ernest Spitzer
Marcel L. Geleijnse
Nahid El Faquir
Cardiology
Amsterdam Cardiovascular Sciences
ACS - Heart failure & arrhythmias
Source :
European Heart Journal-Cardiovascular Imaging, 19(2), 157-167. Oxford University Press, European heart journal cardiovascular Imaging, 19(2), 157-167. Oxford University Press
Publication Year :
2018

Abstract

Aims The Lotus and SAPIEN3 are second-generation transcatheter heart valves, which are designed to minimize paravalvular aortic regurgitation (PAR) after transcatheter aortic valve replacement (TAVR). We sought to compare both devices for valve performance and with emphasis on PAR by independent core laboratory analysis. Methods and results A total of 162 (79 Lotus and 83 SAPIEN3) consecutive patients (51% female, 80 ± 8 years, Logistic EuroSCORE 14.8 ± 9.4%) who underwent TAVR because of aortic stenosis were included. Patients with aortic valve-in-valve treatment were excluded. Pre-discharge echocardiograms were analysed by an independent core laboratory using the Valve Academic Research Consortium 2 criteria. There were no differences in baseline and procedural characteristics, except for a larger aortic annulus and sizing indices in SAPIEN3-treated patients and frequency of post-dilatation (0% in Lotus and 13.1% in SAPIEN3). Both valves have similar mean residual gradient, indexed effective orifice area and Doppler velocity index when adjusted to valve size. The frequency of mild (13.9% vs. 31.3%) and at least moderate (1.3% vs. 3.6%) PAR was less after Lotus than after SAPIEN3 implantation ( P = 0.02). Multi-slice computed tomography-based annulus and left ventricular outflow tract diameters, calcification and percentage of oversizing were not different between those with or without mild or more PAR. On multivariate analysis, the use of Lotus valve was associated with less (odds ratio OR, 0.41, P = 0.03) occurrence of PAR. Conclusion Overall, haemodynamic performance was comparable between the Lotus and SAPIEN3 valves. Lotus valve required less oversizing and was associated with less PAR than SAPIEN3.

Details

Language :
English
ISSN :
20472404
Database :
OpenAIRE
Journal :
European Heart Journal-Cardiovascular Imaging, 19(2), 157-167. Oxford University Press, European heart journal cardiovascular Imaging, 19(2), 157-167. Oxford University Press
Accession number :
edsair.doi.dedup.....3197e0b61e4e0f88df70cc7a226f32a7