Back to Search
Start Over
A traceable precision current source for currents between 100 aA and 10 pA
- Source :
- Conference Digest Conference on Precision Electromagnetic Measurements.
- Publication Year :
- 2003
- Publisher :
- IEEE, 2003.
-
Abstract
- A new calibration setup for picoamperemeters is presented. Owing to the use of a novel voltage ramp generator with increased linearity, the useful measuring time is increased to about 35 min, which reduces the calibration uncertainty of picomaperemeters with long time constants.
Details
- Database :
- OpenAIRE
- Journal :
- Conference Digest Conference on Precision Electromagnetic Measurements
- Accession number :
- edsair.doi.dedup.....32721a1b94bd5efa2e47db31d4fbe99f
- Full Text :
- https://doi.org/10.1109/cpem.2002.1034767