Increased capture threshold in permanent His‐bundle pacing associated with flecainide

Authors :
Bradley P. Knight
Jeremiah Wasserlauf
Nishant Verma
Michael Jiang
Source :
Pacing and Clinical Electrophysiology. 43:360-363
Publication Year :
2020
Publisher :
Wiley, 2020.

Abstract

A 64-year-old man underwent implantation of a permanent His-bundle pacemaker. A marked rise in the selective His-bundle capture threshold was noted 1 month after the patient started flecainide acetate for rhythm control of recurrent, symptomatic atrial flutter and atrial fibrillation. The capture threshold subsequently normalized 4 days after discontinuing flecainide and switching to dofetilide. To our knowledge, this is the first documented case of a rise in selective His-bundle capture threshold associated with flecainide acetate. Further studies are needed to characterize this association which could result in higher capture thresholds, decreased battery longevity, and mimic His-bundle lead failure.