Back to Search Start Over

Comparison of tracheal tube cuff diameters with internal transverse diameters of the trachea in children

Authors :
Maja I. Hug
Felix Berger
Alexander Dullenkopf
Walter Knirsch
K. Stutz
Markus Weiss
Oliver Kretschmar
Maren Tomaske
Source :
Acta Anaesthesiologica Scandinavica. 50:201-205
Publication Year :
2006
Publisher :
Wiley, 2006.

Abstract

Background: In this study, we evaluated the ratio of the cuff diameters of the Microcuff paediatric tracheal tube (Microcuff PET, Weinheim, Germany) to fluoroscopically measured internal transverse tracheal diameters in children from birth to adolescence. Methods: With Institutional Ethics Committee approval and parental consent, we measured the internal transverse tracheal diameters from fluoroscopy images in children undergoing cardiac catheterization requiring general anaesthesia with oro-tracheal intubation. Minimal tracheal sealing pressures were assessed at standardized respirator settings. Internal transverse tracheal diameters were compared with cuff diameters at 20 cmH2O cuff pressure. Linear regression analysis was employed to assess the correlation of tracheal diameters with age, height and weight, and to assess the correlation of the cuff/tracheal diameter ratio with sealing pressures. For all tests, P

Details

ISSN :
13996576 and 00015172
Volume :
50
Database :
OpenAIRE
Journal :
Acta Anaesthesiologica Scandinavica
Accession number :
edsair.doi.dedup.....330ce8c5d2e8149ae1104d639a8a166d
Full Text :
https://doi.org/10.1111/j.1399-6576.2006.00943.x