- WOS: 000372609000012<br />A novel and highly sensitive sensor based on poly[2, 5-di(2-thiophenyl)-1-p-(tolyl) pyrrole] was developed and successfully applied for the electrochemical determination of acetaminophen (AC) in pharmaceutical samples. The voltammetric behaviors of AC on modified pencil graphite electrode were investigated by cyclic voltammetry and differential pulse voltammetry in phosphate buffer solution. The voltammetric studies reveal that the polymer modified pencil graphite electrode possesses a remarkable electrocatalytic activity toward acetaminophen compared with the unmodified pencil graphite electrode. A linear dynamic range from 0.025 to 5 mu M and a correlation coefficient of 0.997 with a slope (sensitivity) of 23 mA M-1 were obtained for AC. The detection limit (S/N = 3) was determined as 1.94 nM, which shows a high sensitivity of the developed sensor for AC.<br />Anadolu University Research Projects Commission [1401F016]<br />This work was supported by the Anadolu University Research Projects Commission under Grant 1401F016.
- Subjects :
- Detection limit
Materials science
010401 analytical chemistry
Analytical chemistry
Electropolymerization
02 engineering and technology
021001 nanoscience & nanotechnology
Electrochemistry
01 natural sciences
0104 chemical sciences
Substituted Pyrrole
chemistry.chemical_compound
chemistry
Electrode
Voltammetry
Graphite
Differential pulse voltammetry
Pencil Graphite Electrode
Electrical and Electronic Engineering
Cyclic voltammetry
0210 nano-technology
Instrumentation
Pyrrole
Acetaminophen
Subjects
Details
- Language :
- English
- Database :
- OpenAIRE
- Accession number :
- edsair.doi.dedup.....33fe888b565554b2b89591e1dcfb66fe