Image Noise and Radiation Dose Using an Automatic Tube Current Modulation Technique at 64-Detector Computed Tomography
- Source :
- Journal of Computer Assisted Tomography. 33:973-977
- Publication Year :
- 2009
- Publisher :
- Ovid Technologies (Wolters Kluwer Health), 2009.
-
Abstract
- We describe the effect of the off-center position, the bowtie filter type, and the direction of scan projection radiograph on the radiation dose and image noise of 64-detector computed tomographic scans using the automatic tube current modulation technique. On anteroposterior scan projection radiographs, the tube current ratio increased, with an increase in the off-center position of the phantom irrespective of the filter size. On lateral scan projection radiographs, it remained constant with each bowtie filter type regardless of the off-center phantom position.