Multi frequency phase fluorimetry (MFPF) for oxygen partial pressure measurement: ex vivo validation by polarographic clark-type electrode

Authors :
James E. Baumgardner
Jolanda Consiglio
Andreas Vogt
Klaus Markstaller
Reto Basciani
Bastian Duenges
Beate Mayr
Klaus Ulrich Klein
Stefan Boehme
Volker Hartwich
Source :
PLoS ONE, Vol 8, Iss 4, p e60591 (2013), PLoS ONE, Boehme, S; Duenges, B; Klein, KU; Hartwich, Volker; Mayr, B; Consiglio, J; Baumgardner, JE; Markstaller, Klaus Michael; Basciani, Reto Marco; Vogt, Andreas (2013). Multi frequency phase fluorimetry (MFPF) for oxygen partial pressure measurement: ex vivo validation by polarographic clark-type electrode. PLoS ONE, 8(4), e60591. Public Library of Science 10.1371/journal.pone.0060591
Publication Year :
2013
Publisher :
Public Library of Science (PLoS), 2013.

Abstract

BACKGROUND: Measurement of partial pressure of oxygen (PO2) at high temporal resolution remains a technological challenge. This study introduces a novel PO2 sensing technology based on Multi-Frequency Phase Fluorimetry (MFPF). The aim was to validate MFPF against polarographic Clark-type electrode (CTE) PO2 measurements. METHODOLOGY/PRINCIPAL FINDINGS: MFPF technology was first investigated in N = 8 anaesthetised pigs at FIO2 of 0.21, 0.4, 0.6, 0.8 and 1.0. At each FIO2 level, blood samples were withdrawn and PO2 was measured in vitro with MFPF using two FOXY-AL300 probes immediately followed by CTE measurement. Secondly, MFPF-PO2 readings were compared to CTE in an artificial circulatory setup (human packed red blood cells, haematocrit of 30%). The impacts of temperature (20, 30, 40°C) and blood flow (0.8, 1.6, 2.4, 3.2, 4.0 L min(-1)) on MFPF-PO2 measurements were assessed. MFPF response time in the gas- and blood-phase was determined. Porcine MFPF-PO2 ranged from 63 to 749 mmHg; the corresponding CTE samples from 43 to 712 mmHg. Linear regression: CTE = 15.59+1.18*MFPF (R(2) = 0.93; P0.05). MFPF response-time (monoexponential) was 1.48±0.26 s for the gas-phase and 1.51±0.20 s for the blood-phase. CONCLUSIONS/SIGNIFICANCE: MFPF-derived PO2 readings were reproducible and showed excellent correlation and good agreement with Clark-type electrode-based PO2 measurements. There was no relevant impact of temperature and blood flow upon MFPF-PO2 measurements. The response time of the MFPF FOXY-AL300 probe was adequate for real-time sensing in the blood phase.

Details

Language :
English
ISSN :
19326203
Volume :
8
Issue :
4
Database :
OpenAIRE
Journal :
PLoS ONE
Accession number :
edsair.doi.dedup.....384a840632927bdcaa32394c4bc88bc7
Full Text :
https://doi.org/10.1371/journal.pone.0060591