Back to Search Start Over

A micro-bioimpedance meter for monitoring insulin bioavailability in personalized diabetes therapy

Authors :
Umberto Cesaro
Maurizio Taglialatela
Nicola Moccaldi
Pasquale Arpaia
Mirco Frosolone
Arpaia, P.
Cesaro, U.
Frosolone, M.
Moccaldi, N.
Taglialatela, M.
Source :
Scientific Reports, Scientific Reports, Vol 10, Iss 1, Pp 1-11 (2020)
Publication Year :
2020

Abstract

An on-chip transducer, for monitoring noninvasively the insulin bio-availability in real time after administration in clinical diabetology, is proposed. The bioavailability is assessed as insulin decrease in situ after administration by means of local impedance measurement. Inter-and-intra individual reproducibility is enhanced by a personalized model, specific for the subject, identified and validated during each insulin administration. Such a real-time noninvasive bioavailability measurement allows to increase the accuracy of insulin bolus administration, by attenuating drawbacks of glycemic swings significantly. In the first part of this paper, the concept, the architecture, and the operation of the transducer, as well as details about its prototype, are illustrated. Then, the metrological characterization and validation are reported in laboratory, in vitro on eggplants, ex vivo on pig abdominal non-perfused muscle, and in vivo on a human subject, using injection as a reference subcutaneous delivery of insulin. Results of significant intra-individual reproducibility in vitro and ex vivo point out noteworthy scenarios for assessing insulin bioavailability in clinical diabetology.

Details

Language :
English
Database :
OpenAIRE
Journal :
Scientific Reports, Scientific Reports, Vol 10, Iss 1, Pp 1-11 (2020)
Accession number :
edsair.doi.dedup.....396f53bd23e946020ae8ec15e6dbeccc