Discrepant post filter ionized calcium concentrations by common blood gas analyzers in CRRT using regional citrate anticoagulation
- Source :
- Critical Care
- Publication Year :
- 2015
- Publisher :
- BioMed Central, 2015.
-
Abstract
- Introduction Ionized calcium (iCa) concentration is often used in critical care and measured using blood gas analyzers at the point of care. Controlling and adjusting regional citrate anticoagulation (RCA) for continuous renal replacement therapy (CRRT) involves measuring the iCa concentration in two samples: systemic with physiological iCa concentrations and post filter samples with very low iCa concentrations. However, modern blood gas analyzers are optimized for physiological iCa concentrations which might make them less suitable for measuring low iCa in blood with a high concentration of citrate. We present results of iCa measurements from six different blood gas analyzers and the impact on clinical decisions based on the recommendations of the dialysis’ device manufacturer. Method The iCa concentrations of systemic and post filter samples were measured using six distinct, frequently used blood gas analyzers. We obtained iCa results of 74 systemic and 84 post filter samples from patients undergoing RCA for CRRT at the University Medicine of Greifswald. Results The systemic samples showed concordant results on all analyzers with median iCa concentrations ranging from 1.07 to 1.16 mmol/L. The medians of iCa concentrations for post filter samples ranged from 0.21 to 0.50 mmol/L. Results of >70 % of the post filter samples would lead to major differences in decisions regarding citrate flow depending on the instrument used. Conclusion Measurements of iCa in post filter samples may give misleading information in monitoring the RCA. Recommendations of the dialysis manufacturer need to be revised. Meanwhile, little weight should be given to post filter iCa. Reference methods for low iCa in whole blood containing citrate should be established.
- Subjects :
- medicine.medical_specialty
Letter
medicine.medical_treatment
Critical Care and Intensive Care Medicine
Internal medicine
Hemofiltration
medicine
Citrate anticoagulation
Humans
False Positive Reactions
Renal replacement therapy
Citrates
False Negative Reactions
Dialysis
Point of care
Whole blood
Calcium metabolism
business.industry
Research
Reproducibility of Results
Anticoagulants
Surgery
Filter (video)
Cardiology
cardiovascular system
Calcium
Blood Gas Analysis
business
Subjects
Details
- Language :
- English
- ISSN :
- 1466609X and 13648535
- Volume :
- 19
- Issue :
- 1
- Database :
- OpenAIRE
- Journal :
- Critical Care
- Accession number :
- edsair.doi.dedup.....3a38facea0a9d7ae90fdece84462dff5