Back to Search Start Over

Permanent pacemaker implantation in premature infants less than 2,000 grams of body weight

Authors :
Hitoshi Mohri
Kiyoshi Haneda
Motohisa Tofukuji
Kaori Sato
Takahiko Nakame
Naoshi Sato
Mikio Ohmi
Source :
The Annals of Thoracic Surgery. 54:1223-1225
Publication Year :
1992
Publisher :
Elsevier BV, 1992.

Abstract

Pacemaker implantation in premature infants presents technical problems because of the relatively larger size of the pulse generator compared with their bodies. A new technique with which successful generator implantation was performed in 2 premature infants less than 2,000 g of body weight is described. The generator is wrapped in a Gore-Tex surgical membrane. A piece of membrane overlying the electrical contact surface of the generator is removed, and the generator is fixed to the abdominal wall in the peritoneal cavity. The technique is simple to perform and would give relative ease in generator exchange.