Back to Search Start Over

Estimating regional pulmonary blood flow in EIT with regularized deconvolution with a Tikhonov regularization

Authors :
Birgit Stender
Marcelo Gama de Abreu
Olaf Dössel
Xenia Augustin
Thomas Bluth
Michael Kircher
Source :
Current Directions in Biomedical Engineering, Vol 6, Iss 3, Pp 60-63 (2020)
Publication Year :
2020
Publisher :
De Gruyter, 2020.

Abstract

Electrical Impedance Tomography (EIT) is a clinically used tool for bed-side monitoring of ventilation. Previous work also showed a high potential for lung perfusion monitoring with indicator-enhanced EIT. However, many research questions have yet to be answered before it can be broadly applied in clinical everyday life. The goal of this work is to evaluate a new method to improve EIT perfusion measurements. Pulmonary hemodynamic transfer functions were estimated using regularized deconvolution with Tikhonov regularization to estimate spatial perfusion parameters. The final comparison between EIT images and PET scans showed a median correlation of 0.897 for the images which were reconstructed using the regularized deconvolution. In comparison the previously used maximum slope method led to a median correlation of 0.868.