Back to Search Start Over

What is the underestimation of radiation dose to the pediatric thyroid gland from contrast enhanced CT, if contrast medium uptake is not taken into account?

Authors :
Adam Hatzidakis
John Damilakis
Maria Raissaki
Antonis Tzedakis
Kostas Perisinakis
Kostas Spanakis
Styliani Pouli
Source :
Physica Medica. 49:95-98
Publication Year :
2018
Publisher :
Elsevier BV, 2018.

Abstract

Purpose To assess the underestimation of radiation dose to the thyroid of children undergoing contrast enhanced CT if contrast medium uptake is not taken into account. Methods 161 pediatric head, head & neck and chest CT examinations were retrospectively studied to identify those involving pre- and post-contrast imaging and thyroid inclusion in imaged volume. CT density of thyroid tissue in HU was measured in non-enhanced (NECT) and corresponding contrast-enhanced CT (CECT) images. Resulting CT number increase (ΔHU) was recorded for each patient and corresponded to a % w/w iodine concentration. The relation of %w/w iodine concentration to %dose increase induced by iodinated contrast uptake was derived by Monte Carlo simulation experiments. Results The thyroid gland was visible in 11 chest and 3 neck CT examinations involving both pre- and post-contrast imaging. The %w/w concentration of iodine in the thyroid tissue at the time of CECT acquisition was found to be 0.13%–0.58% w/w (mean = 0.26%). The %increase of dose to thyroid tissue was found to be linearly correlated to%w/w iodine uptake. The increase in radiation dose to thyroid due to contrast uptake ranged from 12% to 44%, with a mean value of 23%. Conclusions The radiation dose to the pediatric thyroid from CECT exposure may be underestimated by up to 44% if contrast medium uptake is not taken into account. Meticulous demarcation of imaged volume in pediatric chest CT examinations is imperative to avoid unnecessary direct exposure of thyroid, especially in CT examinations following intravenous administration of contrast medium.

Details

ISSN :
11201797
Volume :
49
Database :
OpenAIRE
Journal :
Physica Medica
Accession number :
edsair.doi.dedup.....41290e9a5570eac3113ef95a9274e0df
Full Text :
https://doi.org/10.1016/j.ejmp.2018.05.009