Back to Search Start Over

MCP-based detectors: calibration and first photon radiation measurements

Authors :
Oleg Brovko
Evgeny Syresin
Wolfgang Freund
Alexander Grebentsov
Mikhail Yurkov
Jan GrĂ¼nert
Source :
Journal of Synchrotron Radiation. 26:1400-1405
Publication Year :
2019
Publisher :
International Union of Crystallography (IUCr), 2019.

Abstract

Detectors based on microchannel plates (MCPs) are used to detect radiation from free-electron lasers. Three MCP detectors have been developed by JINR for the European XFEL (SASE1, SASE2 and SASE3 lines). These detectors are designed to operate in a wide dynamic range from the level of spontaneous emission to the SASE saturation level (between a few nJ up to 25 mJ), in a wide wavelength range from 0.05 nm to 0.4 nm for SASE1 and SASE2, and from 0.4 nm to 4.43 nm for SASE3. The detectors measure photon pulse energies with an anode and a photodiode. The photon beam image is observed with an MCP imager with a phosphor screen. At present, the SASE1 and SASE3 MCP detectors are commissioned with XFEL beams. Calibration and first measurements of photon radiation in multibunch mode are performed with the SASE1 and SASE3 MCPs. The MCP detector for SASE2 and its electronics are installed in the XFEL tunnel, technically commissioned, and are now ready for acceptance tests with the X-ray beam.

Details

ISSN :
16005775
Volume :
26
Database :
OpenAIRE
Journal :
Journal of Synchrotron Radiation
Accession number :
edsair.doi.dedup.....4916cb6ef07891dba1b1e967125b930e
Full Text :
https://doi.org/10.1107/s1600577519006295