Computed Tomographic Colonography
- Source :
- Journal of Computer Assisted Tomography. 32:504-510
- Publication Year :
- 2008
- Publisher :
- Ovid Technologies (Wolters Kluwer Health), 2008.
-
Abstract
- Objective We evaluated an automated polyp size measurement tool in computed tomographic colonography for its accuracy and value for patient risk stratification. Methods A simulation program generated a raw data phantom with sessile and pedunculated polyps of known sizes using 120 to 140 kV and 50, 40, 20, 15, and 10 mAs. All polyps were measured by clicking on the polyp surface. Comparison of the calculated size with the known polyp sizes allowed calculation of reproducibility and accuracy. For patients with proven polyps, we also compared automated measurements with manual and endoscopic measurements to evaluate the effect on patient risk stratification. Results The automated measurement tool allowed accurate measurements. In the patient study, assignment to the correct size group was not significantly different from the radiologist's results. However, it slightly improved patient risk stratification by reducing both failed and unnecessary colonoscopy referral. Conclusions An automated tool for polyp measurement in patients facilitates patient risk stratification.
- Subjects :
- Adult
Male
medicine.medical_specialty
Software Validation
Patient risk
Colonic Polyps
Colonoscopy
Risk Assessment
Severity of Illness Index
Imaging phantom
Outpatients
Image Processing, Computer-Assisted
otorhinolaryngologic diseases
medicine
Humans
Radiology, Nuclear Medicine and imaging
Computed Tomographic Colonography
In patient
Aged
Observer Variation
Reproducibility
medicine.diagnostic_test
Phantoms, Imaging
business.industry
Reproducibility of Results
Polyp size
Middle Aged
digestive system diseases
surgical procedures, operative
Feasibility Studies
Female
Radiology
business
Risk assessment
Colonography, Computed Tomographic
Software
Subjects
Details
- ISSN :
- 03638715
- Volume :
- 32
- Database :
- OpenAIRE
- Journal :
- Journal of Computer Assisted Tomography
- Accession number :
- edsair.doi.dedup.....4947c87bfbc1d5c2000899c0fd36b01c
- Full Text :