Evaluation of optical performance of 4 aspheric toric intraocular lenses using an optical bench system: Influence of pupil size, decentration, and rotation
- Source :
- Journal of cataract and refractive surgery. 41(10)
- Publication Year :
- 2014
-
Abstract
- Purpose To evaluate the effect of pupil size, degree of intraocular lens (IOL) decentration, and rotation of 4 aspheric toric IOLs on the image quality. Setting Department of Ophthalmology, Seoul St. Mary's Hospital, Seoul, South Korea. Design Experimental study. Methods Four aspheric toric intraocular lenses (IOLs)—the Precizon (transitional conic toric IOL), AT Torbi 709M (bitoric IOL), SN6AT4 (posterior toric surface IOL), and ZCT225 (anterior toric surface IOL)—were evaluated using the optical bench metrology system. Measurements included changes in spherical aberrations, relative spherical equivalent (SE), and image quality at different pupil diameters and image quality degradation due to decentration and rotation of the IOLs. Results Change in relative SE with pupil size in aberration-free toric IOLs (transitional conic toric and bitoric IOLs) was greater than in negatively aspheric toric IOLs (posterior toric surface and anterior toric surface IOLs). In contrast, the aberration-free IOLs showed higher contrast than the negatively aspheric IOLs. When IOLs were decentered by 1.0 mm, the contrast reduction rates at 17.6 cycles per degree for the transitional conic toric IOL, bitoric IOL, posterior toric surface IOL, and anterior toric surface IOL were 5.1%, 3.1%, 12.2%, and 15.8%, respectively. Rotation-induced deterioration of contrast to 0.5 required a much higher rotation for the transitional conic toric IOL than for the other 3 IOLs. Conclusions The transitional conic toric IOL and bitoric IOL provided superior image quality despite pupil size changes and the presence of decentration. The transitional conic toric IOL demonstrated maximum rotation tolerance compared with the other IOLs. Financial Disclosure No author has a financial or proprietary interest in any material or method mentioned.
- Subjects :
- Lenses, Intraocular
Image quality degradation
Optics and Photonics
Rotation
medicine.medical_treatment
Pupil size
Visual Acuity
Intraocular lens
Spherical equivalent
Pupil
Models, Theoretical
Artificial Lens Implant Migration
Refraction, Ocular
Sensory Systems
Cycles per degree
Ophthalmology
Intraocular lenses
Lens Implantation, Intraocular
medicine
Optometry
Humans
Surgery
Mathematics
Subjects
Details
- ISSN :
- 18734502
- Volume :
- 41
- Issue :
- 10
- Database :
- OpenAIRE
- Journal :
- Journal of cataract and refractive surgery
- Accession number :
- edsair.doi.dedup.....4b0871c1f03b31b43b7716e897a4b86b