Back to Search Start Over

Can driven-right-leg circuits increase interference in ECG amplifiers?

Authors :
Ernesto Serrano-Finetti
Ramon Casanella
Ramon Pallas-Areny
Joan Gomez-Clapers
Source :
EMBC
Publication Year :
2012

Abstract

The Driven-Right-Leg (DRL) circuit has been used for about 50 years to reduce interference due to common-mode voltage in biopotential amplifiers in scenarios that range from fixed equipment supplied from power lines to battery-supplied ambulatory monitors, and for systems that use gelled, dry, textile, and capacitive electrodes. However, power-line interference models predict that for isolation amplifiers, currently mandated by safety standards, power-line interference can often couple mostly in differential mode rather than in common mode. In this work we analyze the effect of the DRL circuit in different ECG leads to elucidate its actual effect on power-line interference reduction. It turns out that that the DRL circuit, which effectively reduces common-mode interference, affects differential-mode interference in an unpredictable way and can increase interference.

Details

ISSN :
26940604
Volume :
2011
Database :
OpenAIRE
Journal :
Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference
Accession number :