Complement to standard E2611-09 of the American Society for Testing and Materials [Standard Test Method for Measurement of Normal Incidence Sound Transmission of Acoustical Materials Based on the Transfer Matrix Method (American Society for Testing and Materials, New York, 2009)] is proposed in order to measure normal incidence sound transmission loss of materials in a modified impedance tube using a three-microphone two-load or one-load method. The modified tube is a standard two-microphone impedance tube, where a third microphone is mounted on a movable hard termination. This method is conceptually identical to the four-microphone two-load or one-load method described in the standard; however, it requires fewer transfer functions and one microphone less. The method is validated on (1) symmetrical homogeneous and (2) non-symmetrical non-homogeneous specimens.