Primary standardisation of technetium-99m by liquid scintillation coincidence counting
- Source :
- Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine. 156
- Publication Year :
- 2019
-
Abstract
- Technetium-99m was standardised by the 4π(LS)ce-γ coincidence extrapolation method. Sensitivity of the 4π(LS) channel to two types of radiation, namely conversion electrons and γ-rays, resulted in incorrect activity values being obtained when this was not adequately accounted for. Measurements were more robust when the LS detection efficiency was optimised, and when a γ-window setting was used that monitored the combined LS efficiency for conversion electrons and γ-rays. The primary standard was internationally compared through participation in the BIPM.RI(II)-K4.Tc99m key comparison.
- Subjects :
- Physics
Radiation
Liquid scintillation counting
Extrapolation
Analytical chemistry
010403 inorganic & nuclear chemistry
01 natural sciences
Coincidence
030218 nuclear medicine & medical imaging
0104 chemical sciences
03 medical and health sciences
0302 clinical medicine
Primary standard
Coincidence counting
Technetium-99m
Sensitivity (electronics)