Diagnostic performance of GENEDIA W and ActiveXpress+ COVID-19 antigens tests among symptomatic individuals in Peru and The United Kingdom

Authors :
Palomino-Padilla, Sandra
Finch, Lorna
de Vos, Margaretha
Savage, Helen Rose
Villa-Castillo, Luz
Hayward, Gail
Cook, Eloïse
Escadafal, Camille
Body, Richard
Adams, Emily R.
Ugarte-Gil, Cesar
Cubas Atienzar, Ana Isabel
Source :
PLOS ONE. 18:e0281925
Publication Year :
2023
Publisher :
Public Library of Science (PLoS), 2023.

Abstract

Objectives In order to generate independent performance data regarding accuracy of COVID-19 antigen-based rapid diagnostic tests (Ag-RDTs), prospective diagnostic evaluation studies across multiple sites are required to evaluate their performance in different clinical settings. This report describes the clinical evaluation the GENEDIA W COVID-19 Ag Device (Green Cross Medical Science Corp., Chungbuk, Korea) and the ActiveXpress+ COVID-19 Complete Testing Kit (Edinburgh Genetics Ltd, UK), in two testing sites Peru and the United Kingdom. Methods Nasopharyngeal swabs collected from 456 symptomatic patients at primary points of care in Lima, Peru and 610 symptomatic participants at a COVID-19 Drive-Through testing site in Liverpool, England were analyzed by Ag-RDT and compared to RT-PCR. Analytical evaluation of both Ag-RDTs was assessed using serial dilutions of direct culture supernatant of a clinical SARS-CoV-2 isolate from the B.1.1.7 lineage. Results For GENEDIA brand, the values of overall sensitivity and specificity were 60.4% [95% CI 52.4–67.9%], and 99.2% [95% CI 97.6–99.7%] respectively; and for Active Xpress+ the overall values of sensitivity and specificity were 66.2% [95% CI 54.0–76.5%], and 99.6% [95% CI 97.9–99.9%] respectively. The analytical limit of detection was determined at 5.0 x 102 pfu/ml what equals to approximately 1.0 x 104 gcn/ml for both Ag-RDTs. The UK cohort had lower median Ct values compared to that of Peru during both evaluations. When split by Ct, both Ag-RDTs had optimum sensitivities at Ct Conclusions Whilst the overall clinical sensitivity of the Genedia did not meet WHO minimum performance requirements for rapid immunoassays in either cohort, the ActiveXpress+ did so for the small UK cohort. This study illustrates comparative performance of Ag-RDTs across two global settings and considers the different approaches in evaluation methods.

Details

ISSN :
19326203
Volume :
18
Database :
OpenAIRE
Journal :
PLOS ONE
Accession number :
edsair.doi.dedup.....4f55db1e04f61274c8ee8cc48ad6bac3