Back to Search
Start Over
Standardization of absolute CD4+ lymphocyte counts across laboratories: An evaluation of the ortho CytoronAbsolute flow cytometry system on normal donors
- Source :
- Cytometry. 22:200-210
- Publication Year :
- 1995
- Publisher :
- Wiley, 1995.
-
Abstract
- The Ortho CytoronAbsolute is a flow cytometer designed to provide direct absolute counts of lymphocytes and their subsets from a single instrument. This study was designed to determine the performance of four geographically separated CytoronAbsolute instruments using 24-h-old, shipped, whole blood samples and to compare the results obtained on the CytoronAbsolute to those obtained using combinations of hematology instruments and other flow cytometers. The absolute count feature of the CytoronAbsolutes located at the four sites were cross calibrated and gave across-site coefficients of variation (CVs) of4.0% for absolute count and 8.2% for absolute lymphocyte count. The calibration was stable for at least 2 months. Absolute lymphocyte counts and lymphocyte percentage immunophenotypes were determined on blood from 50 healthy human immunodeficiency virus (HIV)-seronegative donors. There were no significant site-to-site differences (each P.05) in CD3+/CD4+ absolute lymphocyte counts determined on the CytoronAbsolute. In contrast, there was a significant site-to-site difference (P.001) between sites 2 and 3 and sites 3 and 4 in the absolute CD3+/CD4+ lymphocyte counts determined via the conventional method of combining a flow cytometry-derived percentage with a hematology instrument-derived lymphocyte count. There was no significant difference (P = .388) in CD3+/CD4+ lymphocyte percent determinations between the CytoronAbsolute and the FACScan or Profile II flow cytometers used in this study. These results demonstrate that different operators can cross calibrate CytoronAbsolutes for absolute CD3+/CD4+ lymphocyte subset determinations, even over large geographic distances.
- Subjects :
- Chromatography
medicine.diagnostic_test
Lymphocyte
Significant difference
Biophysics
Reproducibility of Results
Cell Biology
Hematology
Biology
Flow Cytometry
Tissue Donors
CD4 Lymphocyte Count
Pathology and Forensic Medicine
Flow cytometry
Endocrinology
medicine.anatomical_structure
Evaluation Studies as Topic
Absolute count
Calibration
Immunology
medicine
Humans
CD4 Lymphocyte
Lymphocyte subsets
Subjects
Details
- ISSN :
- 10970320 and 01964763
- Volume :
- 22
- Database :
- OpenAIRE
- Journal :
- Cytometry
- Accession number :
- edsair.doi.dedup.....50a0600ac7a6fb287e6f30ceee7941be
- Full Text :
- https://doi.org/10.1002/cyto.990220307