Back to Search Start Over

Standardization of absolute CD4+ lymphocyte counts across laboratories: An evaluation of the ortho CytoronAbsolute flow cytometry system on normal donors

Authors :
Janis V. Giorgi
Alan L. Landay
Marilyn D. Knight
Mark C. Connelly
Eileen Page
John W. Parker
Thomas J. Mercolino
Jonathan C. Kagan
Cynthia L. Wilkening
Cathie Spino
Source :
Cytometry. 22:200-210
Publication Year :
1995
Publisher :
Wiley, 1995.

Abstract

The Ortho CytoronAbsolute is a flow cytometer designed to provide direct absolute counts of lymphocytes and their subsets from a single instrument. This study was designed to determine the performance of four geographically separated CytoronAbsolute instruments using 24-h-old, shipped, whole blood samples and to compare the results obtained on the CytoronAbsolute to those obtained using combinations of hematology instruments and other flow cytometers. The absolute count feature of the CytoronAbsolutes located at the four sites were cross calibrated and gave across-site coefficients of variation (CVs) of4.0% for absolute count and 8.2% for absolute lymphocyte count. The calibration was stable for at least 2 months. Absolute lymphocyte counts and lymphocyte percentage immunophenotypes were determined on blood from 50 healthy human immunodeficiency virus (HIV)-seronegative donors. There were no significant site-to-site differences (each P.05) in CD3+/CD4+ absolute lymphocyte counts determined on the CytoronAbsolute. In contrast, there was a significant site-to-site difference (P.001) between sites 2 and 3 and sites 3 and 4 in the absolute CD3+/CD4+ lymphocyte counts determined via the conventional method of combining a flow cytometry-derived percentage with a hematology instrument-derived lymphocyte count. There was no significant difference (P = .388) in CD3+/CD4+ lymphocyte percent determinations between the CytoronAbsolute and the FACScan or Profile II flow cytometers used in this study. These results demonstrate that different operators can cross calibrate CytoronAbsolutes for absolute CD3+/CD4+ lymphocyte subset determinations, even over large geographic distances.

Details

ISSN :
10970320 and 01964763
Volume :
22
Database :
OpenAIRE
Journal :
Cytometry
Accession number :
edsair.doi.dedup.....50a0600ac7a6fb287e6f30ceee7941be
Full Text :
https://doi.org/10.1002/cyto.990220307