Back to Search
Start Over
Patient and healthcare professional satisfaction with a new, high accuracy blood glucose meter with color range indicator and wireless connectivity
- Source :
- Expert Review of Medical Devices. 13:619-626
- Publication Year :
- 2016
- Publisher :
- Informa UK Limited, 2016.
-
Abstract
- Accurate self-monitoring of blood glucose is a key component of effective self-management of glycemic control. The OneTouch VerioFlex(™) (OTVF) blood glucose monitoring system (BGMS) was evaluated for accuracy in a clinical setting. Patients also used OTVF for a 1-wk trial period and reported their level of satisfaction with meter features. In a separate study, healthcare professionals used an on-line simulator of the BGMS and answered questions about its potential utility to their patients. OTVF was accurate over a wide glucose range and met lay user and system accuracy blood glucose standards described in ISO15197:2013 as well as the accuracy requirements to fulfill US FDA expectations for 510(k) clearance of BGMS. Patients and healthcare professionals felt the features of OTVF, which has the capability to connect wirelessly to mobile devices and interact wirelessly with diabetes management software, could provide significant benefits to them or their patients.
- Subjects :
- Adult
Blood Glucose
Male
Adolescent
Health Personnel
Biomedical Engineering
Color
030209 endocrinology & metabolism
Personal Satisfaction
Young Adult
03 medical and health sciences
0302 clinical medicine
Diabetes management
medicine
Humans
030212 general & internal medicine
Wireless connectivity
Aged
Glycemic
Aged, 80 and over
Blood glucose monitoring
medicine.diagnostic_test
Health professionals
business.industry
Blood Glucose Self-Monitoring
Glucose meter
General Medicine
Middle Aged
medicine.disease
Patient Satisfaction
Health Care Surveys
Linear Models
Female
Surgery
Medical emergency
business
Wireless Technology
Mobile device
Biomedical engineering
Subjects
Details
- ISSN :
- 17452422 and 17434440
- Volume :
- 13
- Database :
- OpenAIRE
- Journal :
- Expert Review of Medical Devices
- Accession number :
- edsair.doi.dedup.....54746657c80e01e93acbae3c8f7e6202
- Full Text :
- https://doi.org/10.1080/17434440.2016.1195260