Back to Search Start Over

Fast and sensitive detection of an anthrax biomarker using SERS-based solenoid microfluidic sensor

Authors :
Kiweon Cha
Andrew J. deMello
Jun Ho Jeon
Gi-eun Rhie
Jaebum Choo
Jonghoon Choi
Juhui Ko
Rongke Gao
Source :
Biosensors and Bioelectronics. 72:230-236
Publication Year :
2015
Publisher :
Elsevier BV, 2015.

Abstract

We report the application of a fully automated surface-enhanced Raman scattering (SERS)-based solenoid-embedded microfluidic device to the quantitative and sensitive detection of anthrax biomarker poly-γ-D-glutamic acid (PGA) in solution. Analysis is based on the competitive reaction between PGA and PGA-conjugated gold nanoparticles with anti-PGA-immobilized magnetic beads within a microfluidic environment. Magnetic immunocomplexes are trapped by yoke-type solenoids embedded within the device, and their SERS signals were directly measured and analyzed. To improve the accuracy of measurement process, external standard values for PGA-free serum were also measured through use of a control channel. This additional measurement greatly improves the reliability of the assay by minimizing the influence of extraneous experimental variables. The limit of detection (LOD) of PGA in serum, determined by our SERS-based microfluidic sensor, is estimated to be 100 pg/mL. We believe that the defined method represents a valuable analytical tool for the detection of anthrax-related aqueous samples.

Details

ISSN :
09565663
Volume :
72
Database :
OpenAIRE
Journal :
Biosensors and Bioelectronics
Accession number :
edsair.doi.dedup.....54bee741e12de74742c0ab4c44606cb7