Abstract
- Background: Patient-specific design (PSD) total knee arthroplasty implants are marketed to restore neutral mechanical-axis alignment (MAA) and provide better anatomic fit compared with standard off-the-shelf (OTS) total knee arthroplasty designs. The purpose was to compare the Knee Society scores, radiographic outcomes, and complications of PSD and OTS implants. Methods: Retrospective study analyzing PSD and OTS by a single surgeon. Implant design change in PSD occurred during the period of data collection leading to PSD-1 and PSD-2 subgroups. Radiographic data including MAA, femorotibial angle, coronal-tibial angle, tibial slope and patella-sulcus angle, and complications were analyzed. Minimum follow-up was 2 years or until revision, and patients completed Knee Society scores preoperatively and postoperatively at 3, 6, 12, 24 weeks, and final follow-up. Results: There were 136 patients (154 knees), average age (62.76 +/− 8.4 years), and follow-up (3.1 +/− 1.5 years). The groups included PSD-1 (77 knees), PSD-2 (36 knees), and OTS (41 knees). The PSD-2 group had better Knee Society function scores compared with PSD-1 and OTS at all timepoints except final follow-up. PSD-2 had significantly shorter hospital stay (P = 0.000012) and less hemoglobin drop (P = 0.032) compared with PSD-1 and OTS. No differences were observed in MAA (P = 0.349) or final range of motion (P = 0.629) between the 3 groups. PSD-2 had more normal femorotibial angle, coronal-tibial angle, and tibial slope compared with PSD-1 and OTS. Failures requiring revision were 23% (18/77) PSD-1, 0% PSD-2, and 3% (1/35) OTS. Most common modes of failure were tibial subsidence (56%) and polyethylene locking mechanism failure (22%) in PSD-1. Conclusion: Catastrophic failure was seen in the PSD-1 group with tibial subsidence and polyethylene locking mechanism failure. PSD-2 had better early Knee Society function scores, shorter hospital stay, lower hemoglobin drop, radiographic alignment, and no failures compared with PSD-1 and OTS.
- Subjects :
- musculoskeletal diseases
medicine.medical_specialty
business.industry
Radiography
musculoskeletal, neural, and ocular physiology
Total knee arthroplasty
Retrospective cohort study
Patient specific
Single surgeon
Surgery
nervous system
Catastrophic failure
mental disorders
Off the shelf
Medicine
Orthopedics and Sports Medicine
business
Range of motion
psychological phenomena and processes
Research Article
Subjects
Details
- Language :
- English
- ISSN :
- 24747661
- Volume :
- 3
- Issue :
- 11
- Database :
- OpenAIRE
- Journal :
- JAAOS Global Research & Reviews
- Accession number :
- edsair.doi.dedup.....5819b0eb1b77985ece0b4bb7a77973aa