Calibration of NaI (Tl) cylindrical detector using axially shifted radioactive cylindrical sources
- Source :
- Nuclear Technology and Radiation Protection, Vol 34, Iss 4, Pp 353-360 (2019)
- Publication Year :
- 2019
- Publisher :
- National Library of Serbia, 2019.
-
Abstract
- In this article, the full energy peak efficiency of NaI detector using non-axial cylindrical sources is calculated by using a new efficient theoretical approach. This approach depends on using the efficiency transfer method and analytical calculations of the average path length of a gamma photon inside the source to the detector system. Measured efficiencies made by using 152Eu aqueous radioactive cylindrical sources with volumes 25 ml and 400 ml. Comparing calculated efficiencies to the measured one showed good agreement enabling the validation of this approach.
- Subjects :
- Physics
cylindrical sources
Detector
Gamma photon
Average path length
efficiency transfer method
Computational physics
Nuclear Energy and Engineering
full energy peak efficiency
average path length
Calibration
lcsh:QC770-798
lcsh:Nuclear and particle physics. Atomic energy. Radioactivity
effective solid angle ratio
Safety, Risk, Reliability and Quality
Axial symmetry
Energy (signal processing)