Back to Search
Start Over
A platform-independent method to reduce CT truncation artifacts using discriminative dictionary representations
- Source :
- Med Phys
- Publication Year :
- 2017
- Publisher :
- Wiley, 2017.
-
Abstract
- Purpose When the scan field of view (SFOV) of a CT system is not large enough to enclose the entire cross-section of the patient, or the patient needs to be positioned partially outside the SFOV for certain clinical applications, truncation artifacts often appear in the reconstructed CT images. Many truncation artifact correction methods perform extrapolations of the truncated projection data based on certain a priori assumptions. The purpose of this work was to develop a novel CT truncation artifact reduction method that directly operates on DICOM images. Materials and Methods The blooming of pixel values associated with truncation was modeled using exponential decay functions, and based on this model, a discriminative dictionary was constructed to represent truncation artifacts and non-artifact image information in a mutually exclusive way. The discriminative dictionary consists of a truncation artifact sub-dictionary and a non-artifact sub-dictionary. The truncation artifact sub-dictionary contains 1000 atoms with different decay parameters, while the non-artifact sub-dictionary contains 1000 independent realizations of Gaussian white noise that are exclusive with the artifact features. By sparsely representing an artifact-contaminated CT image with this discriminative dictionary, the image was separated into a truncation artifact-dominated image and a complementary image with reduced truncation artifacts. The artifact-dominated image was then subtracted from the original image with an appropriate weighting coefficient to generate the final image with reduced artifacts. This proposed method was validated via physical phantom studies and retrospective human subject studies. Quantitative image evaluation metrics including the relative root mean square error (rRMSE) and the universal image quality index (UQI) were used to quantify the performance of the algorithm. Results For both phantom and human subject studies, truncation artifacts at the peripheral region of the SFOV were effectively reduced, revealing soft tissue and bony structure once buried in the truncation artifacts. For the phantom study, the proposed method reduced the relative RMSE from 15% (original images) to 11%, and improved the UQI from 0.34 to 0.80. Conclusion A discriminative dictionary representation method was developed to mitigate CT truncation artifacts directly in the DICOM image domain. Both phantom and human subject studies demonstrated that the proposed method can effectively reduce truncation artifacts without access to projection data. This article is protected by copyright. All rights reserved.
- Subjects :
- Truncation
Image quality
01 natural sciences
Article
Imaging phantom
030218 nuclear medicine & medical imaging
03 medical and health sciences
DICOM
0302 clinical medicine
Discriminative model
Statistics
Image Processing, Computer-Assisted
Humans
0101 mathematics
Projection (set theory)
Retrospective Studies
Mathematics
Artifact (error)
Pixel
Phantoms, Imaging
business.industry
010102 general mathematics
Pattern recognition
General Medicine
Artificial intelligence
Artifacts
Tomography, X-Ray Computed
business
Algorithms
Subjects
Details
- ISSN :
- 00942405
- Volume :
- 44
- Database :
- OpenAIRE
- Journal :
- Medical Physics
- Accession number :
- edsair.doi.dedup.....597b03bfcfb5e371396b170bfe3f889c
- Full Text :
- https://doi.org/10.1002/mp.12032