Back to Search
Start Over
Jet flow rate and needle position govern distal airway pressures during low‐frequency jet ventilation
- Source :
- Laryngoscope Investigative Otolaryngology, Vol 6, Iss 2, Pp 244-251 (2021), Laryngoscope Investigative Otolaryngology
- Publication Year :
- 2021
- Publisher :
- Wiley, 2021.
-
Abstract
- Objectives Although jet ventilation is frequently used during surgery for airway stenosis, little is known about distal airway pressures during jet ventilation. The objective of the study is to determine how jet pressure, flow rate, and position of the ventilation needle relate to distal airway pressure magnitude and homogeneity. Methods Two 3D models of the first five generations of the human airway tree were created. One is a duplicate of a human airway from a 15‐year‐old healthy male's computed tomography scan, and the other is an idealized symmetric model of human lung morphometry. Pressure transducers measured fifth‐generation distal airway pressures in both models. A computer‐controlled jet needle positioning system was used to ventilate the lung casts. The effects of jet needle position, jet pressure, and jet flow rate on distal airway pressure and homogeneity were measured. Results Total entrained jet flow rate was the most reliable predictor of distal airway pressure. Pressure supplied to the jet ventilation needle had a positive linear relationship with distal airway pressure; however, this relationship was dependent on the jet needle flow resistance. As the ventilation needle moved closer to the tracheal wall, ventilation homogeneity decreased. Depth into the trachea was positively correlated with sensitivity of the needle to the tracheal wall. Conclusion In this model, total entrained jet flow rate is a more robust predictor of distal airway pressure than jet inlet pressure. More homogeneous ventilation was observed in our model with the ventilation needle centered in the proximal region of the trachea.<br />Xxx.
- Subjects :
- Materials science
Flow (psychology)
lcsh:Surgery
complex mixtures
law.invention
pressure
law
jet ventilation
Needle position
Original Research
Jet (fluid)
Airway pressures
General Medicine
lcsh:RD1-811
respiratory system
equipment and supplies
lcsh:Otorhinolaryngology
Laryngology, Speech and Language Science
Pressure sensor
lcsh:RF1-547
Jet ventilation
homogeneity
flow
Ventilation (architecture)
Airway
human activities
Biomedical engineering
circulatory and respiratory physiology
Subjects
Details
- Language :
- English
- ISSN :
- 23788038
- Volume :
- 6
- Issue :
- 2
- Database :
- OpenAIRE
- Journal :
- Laryngoscope Investigative Otolaryngology
- Accession number :