Abstract

Highlights • Image quality is similar for different body phased-array receive coil setups. • An 18-channel body phased-array receive coil setup achieved good image quality. • 60-channel body phased-array receive coil setup slightly improves SNR in T2W images.<br />Purpose To evaluate the influence of body phased-array (BPA) receive coil setups on signal-to-noise ratio (SNR) and image quality (IQ) in prostate MRI. Methods This prospective study evaluated axial T2-weighted images (T2W-TSE) and DWI of the prostate in ten healthy volunteers with 18-channel (18CH), 30-channel and 60-channel (60CH) BPA receive coil setups. SNR and ADC values were assessed in the peripheral and transition zones (TZ). Two radiologists rated IQ features. Differences in qualitative and quantitative image features between BPA receive coil setups were compared. After correction for multiple comparisons, p-values