Effect of CT Localizer Radiographs on Radiation Dose Associated With Automatic Tube Current Modulation: A Multivendor Study
- Source :
- Current problems in diagnostic radiology. 49(1)
- Publication Year :
- 2018
-
Abstract
- Objectives To assess the influence of the CT localizer radiograph on the automatic tube current modulation system of 7 CT scanners produced by 4 different CT manufacturers. Methods The influence of the localizer orientation, table height, tube current and tube potential values on the radiation dose of the related CT scan were evaluated. Images were acquired by using an anthropomorphic phantom positioned in the CT gantry isocenter as well as from -6 cm to +6 cm vertically to the isocenter. Results Vertical movement of the CT table height affected the radiation dose in all scanners using anterior-posterior or a posterior-anterior localizer orientation albeit differently, depending on the manufacturer; only in 1/7 scanner no influence was observed. The latero-lateral localizer orientation proved to be more effective in limiting the influence of the vertical miscentering in all scanners. Changing localizer's tube voltage influenced the scan radiation dose in scanners produced by two manufacturers, while no significant effect was observed in scanners produced by the other two manufacturers. No significant dose variation was observed in 6/7 scanners when changing the localizer's tube current. Conclusion Localizer radiograph shows a significant influence on the radiation exposure but with different outcomes depending on the manufacturer of the CT scanner. Radiologists and radiographers should have a thorough understanding of these differences to assure patients the best examination in terms of radiation dose and image quality.
- Subjects :
- Scanner
Tomography Scanners, X-Ray Computed
Orientation (computer vision)
business.industry
Image quality
Phantoms, Imaging
Radiography
Radiation dose
Isocenter
Radiation Exposure
Radiation Dosage
030218 nuclear medicine & medical imaging
03 medical and health sciences
0302 clinical medicine
030220 oncology & carcinogenesis
Tube current modulation
Medicine
Radiology, Nuclear Medicine and imaging
Tomography
business
Nuclear medicine
Tomography, X-Ray Computed
Subjects
Details
- ISSN :
- 15356302
- Volume :
- 49
- Issue :
- 1
- Database :
- OpenAIRE
- Journal :
- Current problems in diagnostic radiology
- Accession number :
- edsair.doi.dedup.....5b9261a55e002b8c833f58c692f8a0f5