Back to Search Start Over

A dynamic calibration technique for temperature programmed desorption spectroscopy

Authors :
Michael J. Heben
A. C. Dillon
P. A. Parilla
Thomas Gennett
Katherine E. Hurst
Jeffrey L. Blackburn
Source :
The Review of scientific instruments. 84(2)
Publication Year :
2013

Abstract

A novel, rapid and accurate calibration procedure as a means for quantitative gas desorption measurement by temperature programmed desorption (TPD) spectroscopy is presented. Quantitative measurement beyond the linear regime of the instrument is achieved by associating an instantaneous calibrated molar flow rate of gas to the detector response. This technique is based on fundamental methods, and is independently verified by comparison to the hydrogen desorption capacity of a known standard metal hydride with known stoichiometry. The TPD calibration procedure described here may be used for any pure gas, and the accuracy is demonstrated for the specific case of hydrogen.

Details

ISSN :
10897623
Volume :
84
Issue :
2
Database :
OpenAIRE
Journal :
The Review of scientific instruments
Accession number :
edsair.doi.dedup.....5d8abff90eecd0a180545f41a01422be