Back to Search Start Over

Evaluation of cyclic fatigue resistance of modern Nickel-Titanium rotary instruments with continuous rotation

Authors :
Dina Abdellatif
Giuseppe Pantaleo
Massimo Amato
Andrea Blasi
Roberto Lo Giudice
Alfredo Iandolo
Source :
Giornale Italiano di Endodonzia, Vol 31, Iss 2, Pp 78-82 (2017)
Publication Year :
2017
Publisher :
PAGEPress Publications, 2017.

Abstract

Aim The aim of present study was to compare cyclic fatigue resistance of three modern Ni–Ti instruments used with continuous rotation. Materials and methods For this study 3 groups of rotating instruments with continuous rotation (HyFlex EDM, Twisted File Adaptive, Revo S SU) have been used, each group consisted of 20 files. The various groups were subjected to cyclic fatigue testing through an artificial metal device. A statistical analysis with Kruskal–Wallis test and Mann–Whitney test was performed. Results There were statistically significant differences between the three groups. The HyFlex EDM instruments have a fracture resistance slightly higher than the Twisted file and far higher than Revo S SU. Conclusions Modern Ni–Ti alloys increase resistance of the rotating instruments to cyclic fatigue.

Details

ISSN :
11214171
Volume :
31
Database :
OpenAIRE
Journal :
Giornale Italiano di Endodonzia
Accession number :
edsair.doi.dedup.....612a3b96fa3efc198509d9c20b3cb526
Full Text :
https://doi.org/10.4081/j.gien.2017.23