Low-dose high-pitch CT angiography of the supraaortic arteries using sinogram-affirmed iterative reconstruction

Authors :
Richard Nolz
Christian Loewe
Christina Plank
Florian Wolf
Rüdiger Schernthaner
Dietrich Beitzke
Veronika Schöpf
Sylvia Unterhumer
Michael Weber
Source :
PLoS ONE, Vol 9, Iss 6, p e99832 (2014), PLoS ONE
Publication Year :
2014
Publisher :
Public Library of Science (PLoS), 2014.

Abstract

ObjectiveTo prospectively evaluate image quality and radiation dose using a low-dose computed tomography angiography protocol and iterative image reconstruction for high-pitch dual-source CT-angiography (DSCTA) of the supraaortic arteries.Material and methodsDSCTA was performed in 42 patients, using either 120 kVp tube voltage, 120 mAS tube current, 2.4 pitch and filtered back projection, or 100 kVp tube voltage, 100 mAs tube current, 3.2 pitch, and sinogram affirmed iterative reconstruction. Measurements of vessel attenuation, of the contrast-to-noise ratio (CNR) and the signal-to-noise ratio (SNR) were performed to objectively evaluate image quality. Two readers evaluated subjective image quality and image noise, using a four-point scale. Effective dose was used to compare the differences in radiation dose.ResultsLow-dose protocol application showed significantly higher vessel opacification (p = 0.013), and non-significantly higher CNR and SNR values. There was no difference in the subjective image quality and image noise reading between the protocols. Effective dose was significantly lower using the low-dose protocol (1.29 ± 0.21 mSv vs. 2.92 ± 0.72 mSv; p < 0.001).ConclusionThe combined use of reduced tube voltage, reduced tube current, and iterative reconstruction reduces radiation dose by 55.4% in high-pitch DSCTA of the supraaortic arteries without impairment of image quality.

Details

Language :
English
ISSN :
19326203
Volume :
9
Issue :
6
Database :
OpenAIRE
Journal :
PLoS ONE
Accession number :
edsair.doi.dedup.....613d60c3fecbe503f212e33fd702d0ca