Predictors of atrial mechanical sensing and atrioventricular synchrony with a leadless ventricular pacemaker: Results from the MARVEL 2 Study

Authors :
Larry A. Chinitz
Todd J. Sheldon
Kurt Stromberg
Jens Brock Johansen
Philippe Ritter
Jonathan P. Piccini
Nicole Wood
Mario Pascual
Clemens Steinwender
Lluís Mont
Joseph Y.S. Chan
Christophe Garweg
Rik Willems
Laurence M. Epstein
Venkata Sagi
Vincent E. Splett
Surinder Kaur Khelae
Source :
Garweg, C, Khelae, S K, Steinwender, C, Chan, J Y S, Ritter, P, Johansen, J B, Sagi, V, Epstein, L M, Piccini, J P, Pascual, M, Mont, L, Willems, R, Sheldon, T, Splett, V, Stromberg, K, Wood, N & Chinitz, L 2020, ' Predictors of atrial mechanical sensing and atrioventricular synchrony with a leadless ventricular pacemaker : Results from the MARVEL 2 Study ', Heart Rhythm, vol. 17, no. 12, pp. 2037-2045 . https://doi.org/10.1016/j.hrthm.2020.07.024
Publication Year :
2020
Publisher :
ELSEVIER SCIENCE INC, 2020.

Abstract

BACKGROUND: The MARVEL (Micra Atrial TRacking Using a Ventricular AccELerometer) 2 study assessed the efficacy of atrioventricular (AV) synchronous pacing with a Micra leadless pacemaker. Average atrioventricular synchrony (AVS) was 89.2%. Previously, low amplitude of the Micra-sensed atrial signal (A4) was observed to be a factor of low AVS. OBJECTIVE: The purpose of this study was to identify predictors of A4 amplitude and high AVS. METHODS: We analyzed 64 patients enrolled in MARVEL 2 who had visible P waves on electrocardiogram for assessing A4 amplitude and 40 patients with third-degree AV block for assessing AVS at rest. High AVS was defined as >90% correct atrial-triggered ventricular pacing. The association between clinical factors and echocardiographic parameters with A4 amplitude was investigated using a multivariable model with lasso variable selection. Variables associated with A4 amplitude together with premature ventricular contraction burden, sinus rate, and sinus rate variability (standard deviation of successive differences of P-P intervals [SDSD]) were assessed for association with AVS. RESULTS: In univariate analysis, low A4 amplitude was inversely related to atrial function assessed by E/A ratio and e'/a' ratio, and was directly related to atrial contraction excursion (ACE) and atrial strain (Ɛa) on echocardiography (all P ≤.05). The multivariable lasso regression model found coronary artery bypass graft history, E/A ratio, ACE, and Ɛa were associated with low A4 amplitude. E/A ratio and SDSD were multivariable predictors of high AVS, with >90% probability if E/A

Details

Language :
English
Database :
OpenAIRE
Journal :
Garweg, C, Khelae, S K, Steinwender, C, Chan, J Y S, Ritter, P, Johansen, J B, Sagi, V, Epstein, L M, Piccini, J P, Pascual, M, Mont, L, Willems, R, Sheldon, T, Splett, V, Stromberg, K, Wood, N & Chinitz, L 2020, ' Predictors of atrial mechanical sensing and atrioventricular synchrony with a leadless ventricular pacemaker : Results from the MARVEL 2 Study ', Heart Rhythm, vol. 17, no. 12, pp. 2037-2045 . https://doi.org/10.1016/j.hrthm.2020.07.024
Accession number :
edsair.doi.dedup.....61828dd5790bd597e5ea87e24fa2d94e
Full Text :
https://doi.org/10.1016/j.hrthm.2020.07.024