Back to Search Start Over

Failure of metal-backed patellar arthroplasty. 47 AGC total knees followed for at least 1 year

Authors :
John Bæch
Hakon Kofoed
Source :
Acta orthopaedica Scandinavica. 62(2)
Publication Year :
1991

Abstract

A prospective series of 47 total knee arthroplasties in 44 patients with gonarthrosis were followed for at least 1 year to detect patellar complications. In five knees the metal-backed patellar component failed, in one knee the cement fractured, and in one knee there was a spontaneous fracture of the patella. We regard this failure rate as unacceptable.

Details

ISSN :
00016470
Volume :
62
Issue :
2
Database :
OpenAIRE
Journal :
Acta orthopaedica Scandinavica
Accession number :
edsair.doi.dedup.....623fe83220b45eceb5b56551adce5926