Quantification of linezolid in serum by LC-MS/MS using semi-automated sample preparation and isotope dilution internal standardization

Authors :
Michael Zoller
Barbara Maier
Johannes Zander
Daniel Teupser
Michael Vogeser
Source :
Clinical chemistry and laboratory medicine. 52(3)
Publication Year :
2013

Abstract

Background: Linezolid serum concentrations have been shown to be highly variable in critically ill patients with often sub-therapeutic drug levels regarding minimal inhibitory concentrations for relevant pathogens. Consequently, therapeutic drug monitoring of linezolid must be considered, requiring a reliable and convenient analytical method. We therefore developed and validated an LC-MS/MS method applying isotope dilution internal standardization and on-line solid phase extraction for serum linezolid quantification. Methods: Sample preparation was based on protein precipitation and on-line solid phase extraction with two-dimensional liquid chromatography and column switching. Three-fold deuterated linezolid was used as the internal standard. The method was validated involving two separate LC-MS/MS systems covering the concentration range of 0.13–32 mg/L. The run time was 4 min. Results: Validation revealed good analytical performance, with inaccuracy Conclusions: This LC-MS/MS assay described herein is a convenient, robust and reliable method for linezolid quantification in serum which can be routinely applied using different LC-MS/MS systems. The method can be used for clinical studies and subsequent TDM of linezolid.

Details

ISSN :
14374331
Volume :
52
Issue :
3
Database :
OpenAIRE
Journal :
Clinical chemistry and laboratory medicine
Accession number :
edsair.doi.dedup.....62e818b59bea5c252f6527c5440349b4