Back to Search
Start Over
A Multicenter Evaluation of a Nongel Mechanical Separator Plasma Blood Collection Tube for Testing of Selected Therapeutic Drugs
- Source :
- The Journal of Applied Laboratory Medicine. 5:671-685
- Publication Year :
- 2020
- Publisher :
- Oxford University Press (OUP), 2020.
-
Abstract
- Background Some therapeutic drugs are unstable during sample storage in gel tubes. BD Vacutainer® Barricor™ Plasma Blood Collection Tube with nongel separator was compared with plasma gel tubes, BD Vacutainer PST™, PST II, and BD Vacutainer Serum Tube for acetaminophen, salicylate, digoxin, carbamazepine, phenytoin, valproic acid, and vancomycin during sample storage for up to 7 days. Methods Seven hospital sites enrolled 705 participants who were taking at least one selected drug. The study tubes were collected and tested at initial time (0 h), after 48 h of storage at room temperature and on day 7 (after additional 5 days of refrigerated storage). The performance of BD Barricor tube was evaluated for each drug by comparing BD Barricor samples with samples from the other tubes at 0 h from the same participant; stability was evaluated by comparing test results from the same tube at 0 h, 48 h, and 7 days. Results At 0 h, BD Barricor showed clinically equivalent results for selected therapeutic drugs compared with the other tubes, except phenytoin in BD PST. Phenytoin samples ≥20 µg/mL in BD PST had 10–12% lower values than samples in BD Barricor. During sample storage, all selected drugs remained stable for 7 days in BD Barricor and in serum aliquots. In BD PST, all drugs remained stable except phenytoin and carbamazepine and in BD PST II except for phenytoin. Conclusion The BD Barricor Tube is effective for the collection and storage of plasma blood samples for therapeutic drug monitoring without sample aliquoting.
- Subjects :
- Phenytoin
Blood Specimen Collection
030213 general clinical medicine
medicine.medical_specialty
medicine.diagnostic_test
Digoxin
business.industry
General Medicine
Carbamazepine
Gastroenterology
Acetaminophen
03 medical and health sciences
0302 clinical medicine
Therapeutic drug monitoring
030220 oncology & carcinogenesis
Internal medicine
medicine
Humans
Vancomycin
Drug Monitoring
Blood Collection Tube
business
Vacutainer
medicine.drug
Subjects
Details
- ISSN :
- 24757241 and 25769456
- Volume :
- 5
- Database :
- OpenAIRE
- Journal :
- The Journal of Applied Laboratory Medicine
- Accession number :
- edsair.doi.dedup.....630db019733b3df7effb9b026f03dd40
- Full Text :
- https://doi.org/10.1093/jalm/jfaa033