Back to Search
Start Over
The Impact of Size and Position of a Mechanical Expandable Transcatheter Aortic Valve: Novel Insights Through Computational Modelling and Simulation
- Source :
- Journal of Cardiovascular Translational Research, 12(5), 435-446. Springer New York
- Publication Year :
- 2019
- Publisher :
- Springer New York, 2019.
-
Abstract
- Transcatheter aortic valve implantation has become an established procedure to treat severe aortic stenosis. Correct device sizing/positioning is crucial for optimal outcome. Lotus valve sizing is based upon multiple aortic root dimensions. Hence, it often occurs that two valve sizes can be selected. In this study, patient-specific computer simulation is adopted to evaluate the influence of Lotus size/position on paravalvular aortic regurgitation (AR) and conduction abnormalities, in patients with equivocal aortic root dimensions. First, simulation was performed in 62 patients to validate the model in terms of predicted AR and conduction abnormalities using postoperative echocardiographic, angiographic and ECG-based data. Then, two Lotus sizes were simulated at two positions in patients with equivocal aortic root dimensions. Large valve size and deep position were associated with higher contact pressure, while only large size, not position, significantly reduced the predicted AR. Despite general trends, simulations revealed that optimal device size/position is patient-specific.
- Subjects :
- 0301 basic medicine
Patient-Specific Modeling
Aortic root
3003 Pharmaceutical Science
Pharmaceutical Science
030204 cardiovascular system & hematology
0302 clinical medicine
Risk Factors
Genetics (clinical)
Models, Cardiovascular
Europe
Treatment Outcome
Aortic Valve
Heart Valve Prosthesis
Equivocal aortic root dimensions
Cardiology
Molecular Medicine
Cardiology and Cardiovascular Medicine
Computer simulations
medicine.medical_specialty
2716 Genetics (clinical)
Transcatheter aortic
Aortic Valve Insufficiency
Clinical Decision-Making
610 Medicine & health
Aortic regurgitation
Regurgitation (circulation)
Prosthesis Design
Risk Assessment
2705 Cardiology and Cardiovascular Medicine
Transcatheter Aortic Valve Replacement
TAVI
03 medical and health sciences
1311 Genetics
Position (vector)
Internal medicine
Multidetector Computed Tomography
Genetics
medicine
Humans
In patient
Conduction abnormalities
business.industry
Hemodynamics
Arrhythmias, Cardiac
Aortic Valve Stenosis
medicine.disease
10020 Clinic for Cardiac Surgery
Stenosis
030104 developmental biology
1313 Molecular Medicine
business
Contact pressure
Subjects
Details
- ISSN :
- 19375395 and 19375387
- Volume :
- 12
- Issue :
- 5
- Database :
- OpenAIRE
- Journal :
- Journal of Cardiovascular Translational Research
- Accession number :
- edsair.doi.dedup.....64fe6afea85b5757cf5c002cc48d87b6