Back to Search
Start Over
One-Year Safety and Performance Assessment of the Argus II Retinal Prosthesis: A Postapproval Study
- Source :
- JAMA Ophthalmol
- Publication Year :
- 2019
-
Abstract
- Importance The Argus II Retinal Prosthesis System is indicated for patients with vision loss due to severe to profound outer retinal degeneration, a group with few treatment options. Objectives To collect postapproval safety and visual function data for the Argus II. Design, Setting, and Participants Multicenter, postapproval clinical trial conducted at 9 sites in Germany and Italy. Data were collected from December 2, 2011, to September 30, 2017, and patients were followed-up for 12 months or longer. Patients were 25 years or older with severe to profound outer retinal degeneration, some residual light perception or the ability of the retina to respond to electrical stimulation, and a history of useful form vision and were already planning to undergo Argus II implantation. Main Outcomes and Measures The primary end point of this study was the nature and rate of adverse events. Secondary end points included 3 visual function tests: square localization (SL), direction of motion, and grating visual acuity (GVA). Results Forty-seven patients were followed for 12 months or longer after implant. Mean (SD) age was 56 (12) years, 37 (79%) had retinitis pigmentosa, and 27 (57%) were male. Through the first 12 months postimplantation, 23 patients (49%) experienced 51 nonserious adverse events and 12 (26%) experienced 13 serious adverse events (SAEs), 9 of which were judged to be related to the Argus II, and 4 of which were judged to be related to the procedure. The most common SAE was conjunctival erosion, reported in 4 patients. No significance testing was done for group analysis for the SL or direction-of-motion tests. When averaged across the group, patients’ accuracy on the SL test, but not on the direction-of-motion test, appeared better when the Argus II was on than when it was switched off. For GVA, more patients at each point in time achieved the 2.9 GVA cutoff in the implanted eye when the Argus II was on compared with it switched off. Conclusions and Relevance Safety and visual function outcomes in this clinical practice setting cohort of patients with Argus II implants were consistent with previously reported results. Longer follow-up of these patients and data from additional patients are required to better outline the risks and benefits of this approach to addressing blindness secondary to severe-to-profound outer retinal degeneration. Trial Registration ClinicalTrials.gov identifier:NCT01490827
- Subjects :
- Retinal degeneration
Pediatrics
medicine.medical_specialty
Visual acuity
01 natural sciences
Prosthetics
Visual Prosthesis
Prosthetic vision
NO
03 medical and health sciences
0302 clinical medicine
Retinitis pigmentosa
Clinical endpoint
medicine
0101 mathematics
Adverse effect
Original Investigation
business.industry
Settore MED/30 - MALATTIE APPARATO VISIVO
010102 general mathematics
medicine.disease
Clinical trial
Ophthalmology
Visual prosthesis
Cohort
030221 ophthalmology & optometry
Argus II Retinal Prosthesis System
medicine.symptom
business
Subjects
Details
- Language :
- English
- Database :
- OpenAIRE
- Journal :
- JAMA Ophthalmol
- Accession number :
- edsair.doi.dedup.....65adc2146fbc9f6979550447d0c41b41