Setup error analysis in helical tomotherapy based image-guided radiation therapy treatments
- Source :
- Journal of Medical Physics / Association of Medical Physicists of India, Journal of Medical Physics, Vol 40, Iss 4, Pp 233-239 (2015)
- Publication Year :
- 2016
-
Abstract
- The adequacy of setup margins for various sites in patients treated with helical tomotherapy was investigated. A total of 102 patients were investigated. The breakdown of the patients were as follows: Twenty-five patients each in brain, head and neck (H and N), and pelvis, while 12 patients in lung and 15 in craniospinal irradiation (CSI). Patients were immobilized on the institutional protocol. Altogether 2686 megavoltage computed tomography images were analyzed with 672, 747, 622, 333, and 312 fractions, respectively, from brain, H and N, pelvis, lung, and CSI. Overall systematic and random errors were calculated in three translational and three rotational directions. Setup margins were evaluated using van Herk formula. The calculated margins were compared with the margins in the clinical use for various directions and sites. We found that the clinical isotropic margin of 3 mm was adequate for brain patients. However, in the longitudinal direction it was found to be out of margin by 0.7 mm. In H and N, the calculated margins were well within the isotropic margin of 5 mm which is in clinical use. In pelvis, the calculated margin was within the limits, 8.3 mm versus 10 mm only in longitudinal direction, however, in vertical and lateral directions the calculated margins were out of clinical margins 11 mm versus 10 mm, and 8.7 mm versus 7.0, mm respectively. In lung, all the calculated margins were well within the margins used clinically. In CSI, the variation was found in the middle spine in the longitudinal direction. The clinical margins used in our hospital are adequate enough for sites H and N, lung, and brain, however, for CSI and pelvis the margins were found to be out of clinical margins.
- Subjects :
- lcsh:Medical physics. Medical radiology. Nuclear medicine
medicine.medical_specialty
business.industry
lcsh:R895-920
medicine.medical_treatment
Image guidance
Biophysics
margins
setup error
Tomotherapy
Surgery
Longitudinal direction
medicine.anatomical_structure
Error analysis
Random error
medicine
Technical Note
Radiology, Nuclear Medicine and imaging
In patient
Nuclear medicine
business
Head and neck
Pelvis
Image-guided radiation therapy
Subjects
Details
- ISSN :
- 09716203
- Volume :
- 40
- Issue :
- 4
- Database :
- OpenAIRE
- Journal :
- Journal of medical physics
- Accession number :
- edsair.doi.dedup.....68f6b565e4defb25f58f129870f6ea68