Back to Search Start Over

Uncertainties in alanine dosimetry in the therapeutic dose range

Authors :
Sergey Sholom
Vadim V. Chumak
Vitaly Nagy
Marc F. Desrosiers
Source :
Applied Radiation and Isotopes. 56:917-929
Publication Year :
2002
Publisher :
Elsevier BV, 2002.

Abstract

A method for evaluating the overall uncertainty of alanine EPR transfer dosimetry in the therapeutic dose range is described. The method uses experimental data on EPR signal reproducibility from replicate dosimeters irradiated to low doses (1–5 Gy), estimates of Type B uncertainties, and Monte Carlo simulations of heteroscedastic orthogonal linear regression. A Bruker ECS106 spectrometer and Bruker alanine dosimeters have been used for this evaluation. The results demonstrate that alanine dosimetry can be used for transfer dosimetry in that range with the overall uncertainty 1.5–4% (1σ) depending on the dose, the number of replicate dosimeters, and the duration of the calibration session (the session should not exceed one working day).

Details

ISSN :
09698043
Volume :
56
Database :
OpenAIRE
Journal :
Applied Radiation and Isotopes
Accession number :
edsair.doi.dedup.....6abb3d297244495791fbc34b1f65c728
Full Text :
https://doi.org/10.1016/s0969-8043(01)00271-8