Back to Search Start Over

Comparison of ICare and IOPen vs Goldmann applanation tonometry according to international standards 8612 in glaucoma patients

Authors :
Matthias Kj. Klamann
Milena Pahlitzsch
Eckart Bertelmann
Anna-Karina B. Maier
Necip Torun
Johannes Gonnermann
Jeanette BrĂ¼nner
Source :
International Journal of Ophthalmology, Vol 9, Iss 11, Pp 1624-1628 (2016)
Publication Year :
2016
Publisher :
Press of International Journal of Ophthalmology (IJO PRESS), 2016.

Abstract

AIM: To compare IOPen and ICare rebound tonometry to Goldmann applanation tonometry (GAT) according to International Standards Organization (ISO) 8612 criteria. METHODS: Totally 191 eyes (n=107 individuals) were included. Criteria of ISO 8612 were fulfilled: 3 clusters of IOP, measured by GAT, were formed. The GAT results were given as mean±standard deviation. RESULTS: GAT (19.7±0.5 mm Hg) showed a significant correlation to ICare (19.8±0.5 mm Hg) (r=0.547, P

Details

Language :
English
ISSN :
22274898 and 22223959
Volume :
9
Issue :
11
Database :
OpenAIRE
Journal :
International Journal of Ophthalmology
Accession number :
edsair.doi.dedup.....6dc59f3f17745a3c93cd6bcb68273359