Back to Search Start Over

A New Method of Measuring Effective Energy Using Copper-pipe Absorbers in X-ray CT

Authors :
Kousuke Matsubara
Wataru Mitsui
Tadanori Takata
Hiroji Iida
Tomoyuki Yamamoto
Kimiya Noto
Source :
Japanese Journal of Radiological Technology. 67:1183-1191
Publication Year :
2011
Publisher :
Japanese Society of Radiological Technology, 2011.

Abstract

The general method of measuring the half-value layer (HVL) for X-ray computed tomography (CT) using square aluminum-sheet filters is inconvenient in that the X-ray tube has to be set to stationary mode. To avoid this inconvenience, we investigated a new method using copper-pipe filters that cover the ionization chamber (copper-pipe method). Using this method, the HVL can be measured at the isocenter in the rotation modes of CT. We examined the accuracy and reproducibility of the copper-pipe method compared with those of the general method. The effective energy measured using the copper-pipe method correlated well with the general method (y=1.064x, r=0.987), and its error was 1.81±1.38%. The results indicate that the copper-pipe method enables accurate measurement of the effective energy of X-ray CT and is a convenient method suited to all general X-ray equipment as well as all X-ray CT.

Details

ISSN :
18814883 and 03694305
Volume :
67
Database :
OpenAIRE
Journal :
Japanese Journal of Radiological Technology
Accession number :
edsair.doi.dedup.....72a04806bce103ab930683b35476d842
Full Text :
https://doi.org/10.6009/jjrt.67.1183