In-silico optimisation of tileable philips digital SiPM based thin monolithic scintillator detectors for SPECT applications
- Source :
- Applied Radiation and Isotopes, 168
- Publication Year :
- 2021
- Publisher :
- Elsevier BV, 2021.
-
Abstract
- Over the last decade one of the most significant technological advances made in the field of radiation detectors for nuclear medicine was the development of Silicon Photomultipler (SiPM) sensors. At present only a small number of SiPM based radiation detectors for Single Photon Emission Computed Tomography (SPECT) applications have been explored, and even fewer experimental prototypes developed. An in-silico investigation into the optimal design of a Philips DPC3200 SiPM photosensor-based thin monolithic scintillator detector for SPECT applications was undertaken using the Monte Carlo radiation transport modelling toolkit Geant4 version 10.5. The performance of the 20 different SPECT radiation detector configurations, 4 scintillator materials (NaI(Tl), GAGG(Ce), CsI(Tl) and LaBr3(Ce)) and 5 thicknesses (1–5 mm), were determined through the use of seven figures of merit. It was found that a crystal thickness range of 4–5 mm was required for all four materials to ensure acceptable energy resolution, sensitivity and spatial resolution performance with the Philips DPC3200 SiPM. Any thinner than this and the performance of all four materials was found to degrade rapidly due to a high probability of material specific fluorescence x-ray escape after incident gamma/x-ray photoelectric absorption. When factoring in each material's magnetic resonance imaging compatibility, hygroscopy, and cost, it was found that CsI(Tl) represents the most promising material to construct tileable Philips digital SiPM based thin monolithic scintillator detectors for SPECT applications.
- Subjects :
- Physics - Instrumentation and Detectors
Materials science
Monte Carlo method
FOS: Physical sciences
Geant4
Photodetector
Scintillator
010403 inorganic & nuclear chemistry
01 natural sciences
Particle detector
030218 nuclear medicine & medical imaging
03 medical and health sciences
0302 clinical medicine
Silicon photomultiplier
Figure of merit
Image resolution
Radiation instrumentation
Gamma-ray detector
Radiation
business.industry
Detector
Instrumentation and Detectors (physics.ins-det)
Physics - Medical Physics
0104 chemical sciences
SPECT
Optoelectronics
Medical Physics (physics.med-ph)
business
SPECT/MR
Subjects
Details
- ISSN :
- 09698043
- Volume :
- 168
- Database :
- OpenAIRE
- Journal :
- Applied Radiation and Isotopes
- Accession number :
- edsair.doi.dedup.....75311205d6a169797b63abbf6d46cb56
- Full Text :
- https://doi.org/10.1016/j.apradiso.2020.109368