Back to Search Start Over

Forward treatment planning techniques to reduce the normalization effect in Gamma Knife radiosurgery

Authors :
Hao Wen Cheng
Chun Yuan Kuo
Jo Ting Tsai
Yu-Kai Su
Yu Jen Wang
David Hung-Chi Pan
Wei Lun Lo
Jia Wei Lin
Source :
Journal of Applied Clinical Medical Physics
Publication Year :
2016

Abstract

In Gamma Knife forward treatment planning, normalization effect may be observed when multiple shots are used for treating large lesions. This effect can reduce the proportion of coverage of high‐value isodose lines within targets. The aim of this study was to evaluate the performance of forward treatment planning techniques using the Leksell Gamma Knife for the normalization effect reduction. We adjusted the shot positions and weightings to optimize the dose distribution and reduce the overlap of high‐value isodose lines from each shot, thereby mitigating the normalization effect during treatment planning. The new collimation system, Leksell Gamma Knife Perfexion, which contains eight movable sectors, provides an additional means to reduce the normalization effect by using composite shots. We propose different techniques in forward treatment planning that can reduce the normalization effect. Reducing the normalization effect increases the coverage proportion of higher isodose lines within targets, making the high‐dose region within targets more uniform and increasing the mean dose to targets. Because of the increase in the mean dose to the target after reducing the normalization effect, we can set the prescribed marginal dose at a higher isodose level and reduce the maximum dose, thereby lowering the risk of complications.

Details

ISSN :
15269914
Volume :
18
Issue :
6
Database :
OpenAIRE
Journal :
Journal of applied clinical medical physics
Accession number :
edsair.doi.dedup.....7924841ab9b59edaf7b66a278e757bea