Accurate and Rapid Measurement of Glycated Hemoglobin Using HLC-723 G11 Variant Mode

Authors :
Min-Seung Park
Kyunghoon Lee
Junghan Song
Hyung-Doo Park
Source :
Annals of Laboratory Medicine
Publication Year :
2018

Abstract

Background Type II diabetes mellitus causes many complications, and its prevalence continues to increase in Korea. Accurate measurement of glycated Hb (HbA1c) is important because of its usefulness in diagnosis, follow-up, and prediction of prognosis. We tested the analytical performance of the HLC-723 G11 Variant Mode (G11vr; Tosoh Bioscience, Inc., Tokyo, Japan), recently introduced to Korea, in detecting HbA1c. Methods We evaluated precision, linearity, carry-over, and turnaround time. Using 208 samples, including 108 flagged samples, we compared HbA1c concentrations from four analyzers through correlation analysis: G11vr, HLC-723 G8 Variant Mode (G8vr, Tosoh Bioscience), HLC-723 G11 Standard Mode (G11st, Tosoh Bioscience), and HLC-723 G8 Standard Mode (G8st, Tosoh Bioscience). We used HPLC mass spectrometry (MS) and capillary electrophoresis (CE) to confirm the HbA1c concentrations of 15 additional known Hb variant samples. Results Repeatability (% CV) in measuring low- and high-concentration controls was 0.57% and 0.35%, respectively; within-laboratory precision was 0.86% and 0.69%, respectively. In a linearity test, the coefficient of determination was 0.9999 (measurement range: 3.64% to 18.59%) for HbA1c. The correlations between G11vr and other analyzers were weaker for flagged samples than for non-flagged samples. The carry-over effect was less than 0.4%. Turnaround time for a single sample was lower in G11vr (one minute) than in G8vr (1.6 minutes). For 15 samples with Hb variants, G11vr HbA1c results were more similar than those of other analyzers to HPLC-MS and CE results. Conclusions G11vr showed adequate performance and rapid turnaround time in measuring HbA1c.

Details

ISSN :
22343814
Volume :
39
Issue :
3
Database :
OpenAIRE
Journal :
Annals of laboratory medicine
Accession number :
edsair.doi.dedup.....7940cff4845758184599f754a1f204d1