A multichannel current stimulator chip for spatiotemporal pattern stimulation of neural tissues
- Source :
- EMBC
- Publication Year :
- 2014
- Publisher :
- IEEE, 2014.
-
Abstract
- We developed a prototype very-large-scale integration chip of a multichannel current stimulator for stimulating neural tissues by utilizing 0.25 μm high-voltage complementary metal-oxide-semiconductor technology. Our designed chip has 20 output channels that are driven by five current buffers arranged in parallel; each buffer controls four output channels in time-sharing mode. The amplitude of a stimulation pulse can be controlled within a range of approximately ±100 μA/phase in each output channel. The stimulus parameters, e.g., amplitude and duration, are controlled separately for each channel by digital codes stored in built-in registers. Combinations of anode and cathode electrodes to pass the current can be changed online. We integrated our stimulator chip with a multielectrode array and studied the neuronal responses to multichannel current stimulations with various temporal patterns in mouse brain slices.
- Subjects :
- Materials science
Models, Neurological
Buffers
law.invention
Mice
Spatio-Temporal Analysis
law
Animals
Nerve Tissue
Cerebrum
Electrodes
Visual Cortex
Neurons
Brain
Spatiotemporal pattern
Oxides
Equipment Design
Multielectrode array
Chip
Cathode
Anode
Mice, Inbred C57BL
Amplitude
Semiconductors
Metals
Electrode
Biomedical engineering
Communication channel
Subjects